Safety chambers with vertical airflow

Laminar chamber PCR 1600 series

Adjustable, horizontal laminar air flow of 0.25 to 0.50 m/s or according to guidelines.
Two-stage filtration system: polyamide pre-filter of EU3 (G3) class in accordance with EN 779 and HEPA filter H14.

• The chambers create sterile conditions for the material being processed. It is recommended for working with samples, that do not pose a microbiological threat. They guarantee product protection; they do not guarantee protection of the operator.
• The PCR chamber has 3 germicidal UV lamps, including two with flow sterilization capability air during operation.
• Adjustable, vertical laminar airflow from 0,25 to 0,50
• Dual filtration system: premliminary polyamide EU3 (G3) filter and HEPA H14 with efficiency above 99,995% for MPPS.
Automatic HEPA compensation provides stable and safe volume of the airflow inside the working space.
Silent working fans, CA level according to PN-EN ISO 11201 below 55 dB.
Air velocity control by sensor or volt voltage system with ±10% user correction.
UV bakeriicidal lamp fixed, placed in the upper rear part of the working area protected against accidental switching on during operation.
Hour counterof the working time of the device and the UV lamp of the chamber.
Energy-saving LED lampilluminating the workspace in two light intensity modes.
User-friendly and easy 7” inch touch display „EASYTOUCH” - operated by dedicated buttons with several languages: Polish, English, German, Spanish, French, Italian, Turkish, Greek, Ukrainian.
• Increased comfort of work by launching functions with dedicated buttons (without entering lower menu levels).
• Stainless steel work surface, countertop total as standard.
• Workspace sides with windows.
• Single electrical socket in chamber housing.
• Optional base on lockable wheels.

Technical details

Internal dim. (W x D x H) 1624 x 633 x 780 mm
Internal dim. (W x D x H) 1588 x 625 x 500 mm
Numbers of fans 3
Adjustable airflow 0,25 - 0,50 m/s
Number and type of filters 1 - main HEPA, 1 - prefilter
HEPA filter efficiency >99,995 % dla MPPS - H14
Sound pressure PN-EN ISO 11201 < 55 dB
Lighting intensity > 1400 luxów
Power consumption in lighting operation 62 W
Power consumption in operation without lighting 39 W
Number of UV lamps 3
The power of a germicidal UV lamp 15 W
Number of electrical sockets 2 szt.
Rated power 386 W
Weight of the device with the base 127 kg

PRODUCT CERTIFICATES
• declaration of conformity CE
• the chamber is manufactured in accordance with the certified quality management system ISO 9001

Each chamber is tested in the company's research laboratory operating in accordance with the ISO 17025 standard.

During the test, the following parameters are checked:
• air flow velocity in the working space and barrier,
• number content of particles,
• homogeneity of the filter material (leakage test),
• tightness of the filter mounting,
• deflection of the laminar air stream.

A factory test report is issued and attached to each chamber.

Warranty of the laminar chamber: 24 months.