Safety chambers with vertical airflow
Laminar chamber K 700 PRO series
Safety chambers with vertical airflow
Laminar chamber K 700 PRO series
Adjustable, vertical laminar air flow of 0.25 to 0.50 m/s or according to guidelines.
Two-stage filtration system: polyamide pre-filter of EU3 (G3) class in accordance with EN 779 and HEPA filter H14.
• Automatic filter wear compensation to ensure a constant and safe air flow rate in the chamber.
• Quiet fans, CA level according to PN-EN ISO 11201 below 49 dB.
• Air velocity control by sensor or volt voltage system with ±10% user correction.
• Two filters: pre-filter and main HEPA H14 with efficiency above 99.995% for MPPS.
• UV germicidal lamp with the possibility of programming the working time.
• Hour counter of the working time of the device and the UV lamp of the chamber.
• Energy-saving LED lamp illuminating the workspace.
• Intuitive touch panel control with menus in Polish, English, German, Spanish, Italian, French, Turkish.
• Increased comfort of work by launching functions with dedicated buttons (without entering lower menu levels).
• Stainless steel worktop.
• Workspace sides with windows.
• Single or double electrical socket in chamber housing.
• Optional valves for air, vacuum and gas and base on lockable wheels.
Technical details
Outside dim. (W x H x D) | 728 x 612 x 990 mm |
---|---|
Inside dim. | 694 x 510 x 650 mm |
Number of fans | 1 |
Adjustable airflow | 0,25 - 0,50 m/s |
Number and type of filters | 1 - main HEPA, 1 - pre-filter poliamide |
HEPA filter efficiency | >99,995 % to MPPS - H14 |
Sound pressure PN-EN ISO 11201 | < 49 dB |
Lighting intensity | > 1000 lux |
Power consumption in lighting operation | 33 W |
Power consumption in operation without lighting | 19 W |
Number of UV lamps | 1 in the workspace |
The power of a germicidal UV lamp | 15 W |
Number of electrical sockets | 2 |
Rated power | 201 W |
Weight of the device with the base | 55 kg |
PRODUCT CERTIFICATES
• declaration of conformity CE
• the chamber is manufactured in accordance with the certified quality management system ISO 9001
Each chamber is tested in the company's research laboratory operating in accordance with the ISO 17025 standard.
During the test, the following parameters are checked:
• air flow velocity in the working space and barrier,
• number content of particles,
• homogeneity of the filter material (leakage test),
• tightness of the filter mounting,
• deflection of the laminar air stream.
A factory test report is issued and attached to each chamber.
Warranty of the laminar chamber: 24 months.