Safety chambers with vertical airflow

Laminar chamber PCR 1300 series

Adjustable, horizontal laminar air flow of 0.20 to 0.50 m/s or according to guidelines.
Two-stage filtration system: polyamide pre-filter of EU3 (G3) class in accordance with EN 779 and HEPA filter H14.

Automatic filter wear compensation to ensure a constant and safe air flow rate in the chamber.
• Quiet fans, CA level according to PN-EN ISO 11201 below 49 dB.
Air velocity control by sensor or volt voltage system with ±10% user correction.
Two filters: pre-filter and main HEPA H14 with efficiency above 99.995% for MPPS.
UV bakeriicidal lamp fixed, placed in the upper rear part of the working area protected against accidental switching on during operation.
Two additional UV lamps in the compression chamber to increase the sterility of the air supplied to the workspace. • Hour counterof the working time of the device and the UV lamp of the chamber.
Energy-saving LED lampilluminating the workspace in two light intensity modes.
Intuitive touch panel control with menus in Polish, English, German and Spanish.
• Increased comfort of work by launching functions with dedicated buttons (without entering lower menu levels).
• Stainless steel worktop.
• Workspace sides with windows.
• Single or double electrical socket in chamber housing.
• Optional valves for air, vacuum and gas and base on lockable wheels.

Technical details

Internal dim. (W x H x D) 1319 x 500 x 780 mm
Internal dim. (W x H x D) 1288 x 490 x 500 mm
Numbers of fans 2
Adjustable airflow 0,25 - 0,50 m/s
Number and type of filters 1 - główny HEPA, 1 - wstępny poliamidowy
HEPA filter efficiency >99,995 % dla MPPS - H14
Sound pressure PN-EN ISO 11201 < 54 dB
Lighting intensity > 1350 luxów
Power consumption in lighting operation 84 W
Power consumption in operation without lighting 63 W
Number of UV lamps 2 lampy przepływowe + 1 w przestrzeni roboczej
The power of a germicidal UV lamp 45 W
Number of electrical sockets 1 szt.
Rated power 144 W
Weight of the device with the base 102 kg

PRODUCT CERTIFICATES
• declaration of conformity CE
• the chamber is manufactured in accordance with the certified quality management system ISO 9001

Each chamber is tested in the company's research laboratory operating in accordance with the ISO 17025 standard.

During the test, the following parameters are checked:
• air flow velocity in the working space and barrier,
• number content of particles,
• homogeneity of the filter material (leakage test),
• tightness of the filter mounting,
• deflection of the laminar air stream.

A factory test report is issued and attached to each chamber.

Warranty of the laminar chamber: 36 months.